8980843391e33ccd4285 · Precision Measurement Method for Cryogenic Amplifier Noise Temperatures Below 5 K
Precision Measurement Method for Cryogenic Amplifier Noise Temperatures Below 5 K
We report precision measurements of the effective input noise temperature of a cryogenic (liquid helium temperature) MMIC amplifier at the amplifier reference planes within the cryostat. A method is given for characterizing and removing the effect of the
{
"title_en": "Precision Measurement Method for Cryogenic Amplifier Noise Temperatures Below 5 K",
"designation": "",
"series": "",
"status": "",
"date": "2006-03-01",
"date_original": "March 1, 2006",
"url": "https://www.nist.gov/publications/precision-measurement-method-cryogenic-amplifier-noise-temperatures-below-5-k",
"source": "NIST Publications",
"display_en": "",
"authors_en": "James P. Randa , Eyal Gerecht , Dazhen Gu , Robert L. Billinger",
"abstract_excerpt_en": "We report precision measurements of the effective input noise temperature of a cryogenic (liquid helium temperature) MMIC amplifier at the amplifier reference planes within the cryostat. A method is given for characterizing and removing the effect of the",
"id": "8980843391e33ccd4285",
"sources": [
"NIST Publications"
],
"source_records": [
{
"title_en": "Precision Measurement Method for Cryogenic Amplifier Noise Temperatures Below 5 K",
"designation": "",
"series": "",
"status": "",
"date": "2006-03-01",
"date_original": "March 1, 2006",
"url": "https://www.nist.gov/publications/precision-measurement-method-cryogenic-amplifier-noise-temperatures-below-5-k",
"source": "NIST Publications",
"display_en": "",
"authors_en": "James P. Randa , Eyal Gerecht , Dazhen Gu , Robert L. Billinger",
"abstract_excerpt_en": "We report precision measurements of the effective input noise temperature of a cryogenic (liquid helium temperature) MMIC amplifier at the amplifier reference planes within the cryostat. A method is given for characterizing and removing the effect of the"
}
],